Features
12.1 inch color display with 800×600 resolution
Three working modes: Monitoring, Surgery, Diagnosis
Simple and user-friendly interface
Multi-display modes:
Standard Interface
Large Font
ECG Standard Full Display
OXY
Trend Table
BP Trend
Built-in rechargeable lithium battery (up to 3 hours of continuous use)
VGA output for external display
Technical Specifications
ECG
Lead Mode: 5 Leads (I, II, III, AVR, AVL, AVF, V)
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV
Heart Rate
Adult: 15–300 BPM
Neonatal: 15–350 BPM
Resolution: 1 BPM
Accuracy: ±1%
Sensitivity: µV (Peak-to-peak)
ST Measurement Range: -2.0 to +2.0 mV
ST Accuracy: ±0.02 mV or greater between -0.8 mV to +0.8 mV
Sweep Speed: 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth
Diagnostic: 0.05–130 Hz
Monitoring: 0.5–40 Hz
Surgery: 1–20 Hz
SpO₂
Measuring Range: 0–100%
Resolution: 1%
Accuracy: 70–100% ±2%
Pulse Rate: 20–250 BPM
Pulse Rate Accuracy: ±3 BPM
Resolution: 1 BPM
NIBP (Non-Invasive Blood Pressure)
Method: Oscillometric
Modes: Manual, Auto, STAT
Units: mmHg, kPa
Adult Range
SYS: 40–280 mmHg
DIA: 10–225 mmHg
MEAN: 20–240 mmHg
>Pediatric Range
SYS: 40–220 mmHg
DIA: 10–160 mmHg
MEAN: 20–170 mmHg
>Neonatal Range
SYS: 40–135 mmHg
DIA: 10–100 mmHg
MEAN: 20–110 mmHg
>Resolution: 1 mmHg
>Accuracy: ±5 mmHg
Temperature (TEMP)
Measurement Range: -5 to 50°C
Resolution: 0.1°C
Accuracy: ±0.1°C
Standard Parameters
ECG
RESP (Respiration)
NIBP
TEMP(Temperature)
SpO₂
PR (Pulse Rate)
Optional Parameters
Recorder (Printer)
Central Monitoring System
Dual IBP (Invasive Blood Pressure)
EtCO₂ (End-Tidal CO₂)
12-lead ECG
Nellcor SpO₂
Sun tech NIBP